FDA Adverse Event Other Summary report: N

JAZZ BGMS

MDR report key: 2881400 · Received December 18, 2012

Report

Report Number
3004637226-2012-00043
Event Type
Other
Date Received
December 18, 2012
Date of Event
November 7, 2012
Report Date
December 17, 2012
Manufacturer
AGAMATRIX, INC.
Product Code
NBW
PMA / PMN Number
K072413
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

METER WAS RETURNED AND TESTED PER AGAMATRIX PROCEDURES. NO FAULT WAS FOUND WITH METER. CUSTOMER DID NOT SUFFER HARM DUE TO INCIDENT. THIS CASE HAS BEEN CLOSED.

Description of Event or Problem · 1

PT ATTENDING CLINIC APPOINTMENT - RANDOM BLOOD GLUCOSE ASSAYED ON YSI - RESULT 2.7MMOL/L (3.0 PLASMA EQUIVALENT). PT SURPRISED AS HIS RESULT WAS 6.9 ON JAZZ METER PRIOR TO DRIVING TO CLINIC APPROX 30 MINS AGO. PT ASKED TO BRING METER INTO CLINIC FOR CHECKING. ON (B)(6) 2012: METER COMPARISON DONE BY DIABETES CENTER: YSI RANDOM BLOOD GLUCOSE 6.8 MMOL/L (PLASMA EQUIVALENT 7.6) JAZZ RESULT 10.4MMOL/L. PT TECHNIQUE IN PERFORMING TEST ON METER OBSERVED AND WAS CORRECT. IT WAS INDICATED BY THE DIABETES CENTER DOING THE TESTING THAT NO INJURY WAS SUSTAINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 JAZZ BGMS BLOOD GLUCOSE METER NBW AGAMATRIX, INC. 8000-02345 JM22WS41G

Patients

Seq Age Sex Outcome Treatment
1