FDA Adverse Event
Malfunction
Summary report: N
TOTAL CARE BARIATRIC BED
MDR report key: 2881225
·
Received November 14, 2012
Report
- Report Number
- 1824206-2012-07436
- Event Type
- Malfunction
- Date Received
- November 14, 2012
- Date of Event
- October 24, 2012
- Report Date
- October 24, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN REPLACED THE POWER CORD TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
THE TECHNICIAN ALLEGED THAT THE GROUND RESISTANCE READING WAS OPEN. NO PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTAL CARE BARIATRIC BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1830 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |