FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 2881219 · Received November 14, 2012

Report

Report Number
1314492-2012-00415
Event Type
Malfunction
Date Received
November 14, 2012
Report Date
October 16, 2012
Manufacturer
BAXTER HEALTHCARE CORP.
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SIGMA EVALUATED THE DEVICE IN QUESTION AND THE PUMP-SIDE POLE CLAMP ADAPTOR WAS OBSERVED TO BE DAMAGED. FURTHER EVALUATION FOUND THAT THE PUMP'S AUDIO FUNCTION WAS NOT PRESENT. THE ABSENCE OF AUDIO WAS DUE TO A FAILED 1 WATT SPEAKER. ALL DETECTION CAPABILITIES AND FUNCTIONS PERFORMED AS EXPECTED. THIS REPORT IS BEING SUBMITTED DUE TO THE POTENTIAL FOR PATIENT INJURY THAT COULD RESULT FROM THE LACK OF AUDIO FUNCTION. AT THIS TIME, THE DATE OF EVENT IS UNKNOWN.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP HAD A BROKEN PUMP-SIDE PLE CLAMP ADAPTOR. DURING THE EVALUATION, IT WAS ALSO FOUND THAT THE PUMP HAD NO AUDIO FUNCTION.