LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2012-03010
- Event Type
- Malfunction
- Date Received
- November 16, 2012
- Date of Event
- October 9, 2012
- Report Date
- November 9, 2012
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NURSE
Narratives
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK THERAPY PAD MESSAGES, ADJUST BELT OR CHECK BELT MESSAGES) HAS BEEN CONFIRMED. UPON RECEIPT, THE ELECTRODE BELT FAILED THE TE RECOGNITION TEST. THE CAUSE WAS AN OPEN PULSE WIRE IN THE DISTRIBUTION NODE (DN) TO FRONT TE CABLE. THE CAUSE FOR THE CHECK THERAPY PAD MESSAGES AND ADJUST BELT OR CHECK BELT MESSAGES WAS THE BROKEN PULSE WIRE. THE ROOT CAUSE FOR THE BROKEN PULSE WIRE CANNOT BE POSITIVELY DETERMINED BUT IS LIKELY EXCESSIVE FORCE ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE BROKEN PULSE WIRE. THE PT RECEIVED AN ICD AND THEREFORE, ENDED USE WITH THE DEVICE.
A (B)(6) FEMALE PT'S NURSE CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT CONSTANT CHECK THERAPY PAD AND ADJUST BELT OR CHECK BELT MESSAGES. THE PT RECEIVED AN ICD AND THEREFORE ENDED USE WITH THE DEVICE.