FDA Adverse Event Injury Summary report: N

EON MINI

MDR report key: 2881101 · Received December 14, 2012

Report

Report Number
1627487-2012-03863
Event Type
Injury
Date Received
December 14, 2012
Date of Event
November 19, 2012
Report Date
November 19, 2012
Manufacturer
ST JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-12192011-003-R
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CORRECTION/REMOVAL REPORTING NUMBER: 1627487-05242011-002-R; 1627487-07262012-002-R; 1627487-07262012-001-C. THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD ADVISORY. THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFO RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 1 OF 4. REF MFR REPORTS: 1627487-2012-03864, -03865 AND -03866. THE PT RECEIVED 2 SCS IPGS WITH DIFFERENT SERIAL NUMBERS. THE PT'S SPOUSE REPORTED THE PT IS EXPERIENCING HEATING AT HIS SCS IPG POCKET SITE. THE PT'S SPOUSE ALSO REPORTED THE PT IS EXPERIENCING A SHOCKING SENSATION. ADDITIONALLY, THE PT'S SPOUSE REPORTED THE PT HAS BEEN SCHEDULED TO HAVE BOTH OF HIS SCS SYSTEMS EXPLANTED. ON (B)(6) 2012 ST JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EON MINI SCS IPG LGW ST JUDE MEDICAL - NEUROMODULATION 3788 2887102

Patients

Seq Age Sex Outcome Treatment
1 48 YR Required Intervention SCS LEAD: MODEL 3163| IMPLANT DATE:| IMPLANT DATE:| SCS ANCHOR: MODEL 1194| IMPLANT DATE:| IMPLANT DATE:| SCS LEAD: MODEL 3186| SCS LEAD: MODEL 3166| SCS ANCHOR: MODEL 1192 (2)| IMPLANT DATE: