OT VERIO IQ METER
Report
- Report Number
- 3008382007-2012-08618
- Event Type
- Malfunction
- Date Received
- December 20, 2012
- Report Date
- December 4, 2012
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4): THE TEST STRIPS INVOLVED WITH THIS COMPLAINT WERE EVALUATED AND ER4 WAS OBSERVED WITH CONTROL SOLUTION TESTING. THE METER HAVE ALSO BEEN RETURNED TO LFS; HOWEVER, THE INVESTIGATION HAS NOT BEEN COMPLETED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION HAS BEEN COMPLETED A FOLLOW UP REPORT WILL BE SUBMITTED.
FOLLOW-UP # 2 ¿ (6/24/2013).THE PATIENT¿S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 1/24/2013 AND 6/15/2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS:THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER WAS FOUND TO BE UNABLE TO REPRODUCE AN ERROR 4 MESSAGE. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2012, THE REPORTER CONTACTED LIFESCAN (B)(6), ALLEGING ERROR 4. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3365797 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR |