COULTER® LH 780 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2012-02843
- Event Type
- Malfunction
- Date Received
- December 20, 2012
- Date of Event
- November 23, 2012
- Report Date
- November 23, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K061616
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A BECKMAN COULTER INC. FIELD SERVICE ENGINEER (FSE) VISITED THE SITE AND EVALUATED THE SYSTEM. THE FSE FOUND A LEAKING ASPIRATION NEEDLE. THE FSE REMOVED AND REPLACED THE VENT LINE SENSOR TUBING THAT IS ATTACHED TO THE ASPIRATION NEEDLE. THE ISSUE WAS RESOLVED. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. (B)(4).
THE CUSTOMER REPORTED TO BECKMAN COULTER THAT THE MANUAL PROBE OF THE COULTER LH 780 HEMATOLOGY ANALYZER WAS LEAKING. APPROXIMATELY 6 ML OF DILUENT LEAKED ONTO THE WORKTOP AND THE FLOOR. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT AT THE TIME OF THE EVENT. THERE ARE NO REPORTS OF ANY INJURIES OR EXPOSURES TO ANY LABORATORY PERSONNEL. NO ERRONEOUS RESULTS WERE GENERATED; ACCORDINGLY, THERE WERE NO CHANGES TO PATIENT CARE OR TREATMENT. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER® LH 780 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER, INC. | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |