FDA Adverse Event Malfunction Summary report: N

MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT

MDR report key: 2880663 · Received December 19, 2012

Report

Report Number
2955842-2012-01377
Event Type
Malfunction
Date Received
December 19, 2012
Date of Event
December 10, 2012
Report Date
January 4, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CLOSE CABLE IS BROKEN AT THE DISTAL IDLER PULLEY. THE IDLER PULLEY SPUN FREELY BUT WAS DAMAGED ON THE EDGE AND ON THE SURFACE. THE CABLE SEGMENT STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE NOT DAMAGED.

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS NOT BEEN RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP MDR WILL BE SUBMITTED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE SURGICAL STAFF REPORTED SEEING A BROKEN CABLE ON THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT. NOTHING WAS REPORTED HAVING FALLEN INTO THE PATIENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10120710 505

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES