FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2880575 · Received December 19, 2012

Report

Report Number
2531779-2012-14885
Event Type
Malfunction
Date Received
December 19, 2012
Report Date
December 3, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 (B)(4) 2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP BLACK BOX HISTORY INDICATED MULTIPLE LOSS OF PRIME WARNINGS WITH LOW FORCE BUT SHOWED NO EVIDENCE OF LOSS OF CARTRIDGE DETECTION. THE PUMP PERFORMED THE REWIND AND LOAD STEPS APPROPRIATELY. DURING THE PRIME STEP, THE PUMP HAD A GRINDING NOISE AND THE PUMP WAS UNABLE TO MAINTAIN PRIME. A FORCE SENSOR CALIBRATION TEST WAS PERFORMED AND FOUND THAT THE FORCE SENSOR WAS OUT OF CALIBRATION. THE PUMP WAS OPENED AND FOUND A MISSING BALL BEARING IN THE PUMP DRIVE ASSEMBLY.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS; HOWEVER, THE INVESTIGATION HAS NOT BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION HAS BEEN COMPLETED A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOAD STEP MALFUNCTION) ISSUE AND THE INSULIN WAS DRIPPING FROM THE INFUSION SET TUBING. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 55 YR