FDA Adverse Event
Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 2880538
·
Received December 19, 2012
Report
- Report Number
- 1031452-2012-00457
- Event Type
- Malfunction
- Date Received
- December 19, 2012
- Report Date
- December 19, 2012
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
(B)(4). NO RMA HAS BEEN INITIATED FOR THIS ISSUE. THE MALFUNCTION HAS NOT BEEN CONFIRMED.
Description of Event or Problem · 1
(B)(4). NO SERIOUS INJURY ALLEGED. MALFUNCTION ALLEGED. CONSUMER STATES THAT THE XPO2 WILL PUT OUT OXYGEN FOR ABOUT A MINUTE AND THEN THE UNIT SHUTS COMPLETELY OFF. FOLLOW UP CALL: SPOKE WITH CONSUMER WHO STATED TECH CAME OUT AND LOOKED AT THE UNIT AND SAID THE PROBLEM WAS THEY WERE NOT CHARGING THE UNIT LONG ENOUGH. THERE WAS A SMALL BACKUP PORTABLE UNIT AVAILABLE. NOW THAT THEY ARE CHARGING THE CORRECT TIME EVERYTHING IS WORKING FINE. MDR FILED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) | 868.5440 | CAW | INVACARE FLORIDA OPERATIONS | XPO100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |