FDA Adverse Event Malfunction Summary report: N

COULTER® LH 780 HEMATOLOGY ANALYZER

MDR report key: 2880537 · Received December 19, 2012

Report

Report Number
1061932-2012-02893
Event Type
Malfunction
Date Received
December 19, 2012
Date of Event
November 21, 2012
Report Date
November 21, 2012
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKZ
PMA / PMN Number
K061616
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) DISCOVERED FLUID LEAKED AT PINCH VALVE TUBING, PV46B. THE FSE REPLACED THE LOWER SHEATH RESTRICTOR TUBE AT PV46B AND RESOLVED THE FLUID LEAK ISSUE. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE UNIT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY FIVE MILLILITERS OF FAINT RED FLUID LEAKED OUTSIDE THE INSTRUMENT INVOLVING THE COULTER LH 780 HEMATOLOGY ANALYZER. THE CUSTOMER HAD ON PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LABORATORY COAT AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT IMPACTED. THE CUSTOMER HAS AN EXPOSURE CONTROL/RISK MANAGEMENT PLAN AT THE FACILITY. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COULTER® LH 780 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1