FDA Adverse Event Malfunction Summary report: N

MEGASUTURECUT NEEDLE DRIVER

MDR report key: 2880450 · Received December 19, 2012

Report

Report Number
2955842-2012-01372
Event Type
Malfunction
Date Received
December 19, 2012
Date of Event
November 26, 2012
Report Date
November 26, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CLOSE CABLE WAS FOUND TO BE BROKEN NEAR THE PROXIMAL PULLEYS AND PROXIMAL CLEVIS CABLE HOLE. THE IDLER PULLEY WAS ABLE TO SPIN FREELY AND IT DID NOT EXHIBIT ANY DAMAGE. THE CABLE SEGMENT STICKS OUT AT THE WRIST. THE PROXIMAL CLEVIS CABLE HOLE EXHIBITED WEAR ON ONE EDGE. OTHER CABLES AT WRIST WERE NOT DAMAGED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DAVINCI SI CHOLECYSTECTOMY PROCEDURE, THE CUSTOMER NOTED A 'BROKEN CABLE AT THE DISTAL END' ON THE MEGA SUTURE CUT NEEDLE DRIVER INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEGASUTURECUT NEEDLE DRIVER ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10120620 471

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES