MEGASUTURECUT NEEDLE DRIVER
Report
- Report Number
- 2955842-2012-01372
- Event Type
- Malfunction
- Date Received
- December 19, 2012
- Date of Event
- November 26, 2012
- Report Date
- November 26, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CLOSE CABLE WAS FOUND TO BE BROKEN NEAR THE PROXIMAL PULLEYS AND PROXIMAL CLEVIS CABLE HOLE. THE IDLER PULLEY WAS ABLE TO SPIN FREELY AND IT DID NOT EXHIBIT ANY DAMAGE. THE CABLE SEGMENT STICKS OUT AT THE WRIST. THE PROXIMAL CLEVIS CABLE HOLE EXHIBITED WEAR ON ONE EDGE. OTHER CABLES AT WRIST WERE NOT DAMAGED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DAVINCI SI CHOLECYSTECTOMY PROCEDURE, THE CUSTOMER NOTED A 'BROKEN CABLE AT THE DISTAL END' ON THE MEGA SUTURE CUT NEEDLE DRIVER INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEGASUTURECUT NEEDLE DRIVER | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420309-01 | M10120620 471 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |