FDA Adverse Event
Injury
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2880437
·
Received December 19, 2012
Report
- Report Number
- 2032227-2012-08091
- Event Type
- Injury
- Date Received
- December 19, 2012
- Date of Event
- December 14, 2012
- Report Date
- December 15, 2012
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS. IT WAS STATED THAT THE CUSTOMER HAD A BENT CANNULA, AND THE INSULIN PUMP DID NOT GIVE A NO DELIVERY ALARM. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP PASSED THE HIGH PRESSURE TEST. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-522LCAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |