FDA Adverse Event Malfunction Summary report: N

TENACULUM FORCEPS

MDR report key: 2880433 · Received December 19, 2012

Report

Report Number
2955842-2012-01366
Event Type
Malfunction
Date Received
December 19, 2012
Date of Event
November 21, 2012
Report Date
November 26, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE PITCH UP CABLE WAS FOUND TO BE BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS MISSING IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DAVINCI SI HYSTERECTOMY PROCEDURE, WHILE USING THE TENACULUM FORCEPS INSTRUMENT, CUSTOMER NOTED A 'BROKEN CABLE, NOTHING FELL IN PATIENT OR HARM.'

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TENACULUM FORCEPS ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420207-06 M10100111 340

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES