FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2880274 · Received December 19, 2012

Report

Report Number
1416980-2012-07753
Event Type
Malfunction
Date Received
December 19, 2012
Date of Event
December 3, 2012
Report Date
December 3, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE CUSTOMER REPORTED PROBLEM OF "ALARM F-38" WAS CONFIRMED DURING DEVICE EVALUATION. SERVICE DETERMINED THAT THE CAUSE WAS DUE TO DEFECTIVE FORCE SENSING RESISTORS, WHICH WAS REPLACED TO RESOLVE THE ISSUE. A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A FLOGARD INFUSION PUMP WITH A FAILURE CODE 38. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1