FDA Adverse Event Summary report: N

RELIANCE 444 WASHER/DISINFECTOR

MDR report key: 2880264 · Received December 19, 2012

Report

Report Number
9680353-2012-00125
Date Received
December 19, 2012
Date of Event
November 30, 2012
Report Date
December 19, 2012
Manufacturer
STERIS CANADA CORPORATION
Product Code
MEC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN INSPECTED THE UNIT AND FOUND THE HOSE DRAIN TO BE CRACKED AND LEAKING WATER. THE TECHNICIAN REPLACED THE HOSE, REBUILT THE DRAIN VALVE AND MADE ADJUSTMENTS TO CLAMPS. THE TECHNICIAN RAN A TEST CYCLE AND CONFIRMED THE UNIT TO BE OPERATIONAL.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT WATER WAS LEAKING FROM UNDERNEATH THE RELIANCE 444 WASHER. NO INJURIES OR PROCEDURAL DELAY/CANCELLATIONS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RELIANCE 444 WASHER/DISINFECTOR WASHER/DISINFECTOR MEC STERIS CANADA CORPORATION

Patients

Seq Age Sex Outcome Treatment
1