STERIS SYSTEM 1E
Report
- Report Number
- 3000251274-2012-00273
- Date Received
- December 19, 2012
- Date of Event
- December 6, 2012
- Report Date
- December 19, 2012
- Manufacturer
- STERIS CORPORATION - HOPKINS
- Product Code
- MED
- PMA / PMN Number
- K090036
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A STERIS SERVICE TECHNICIAN ARRIVED ONSITE AND PARTIALLY TURNED ON THE WATER SUPPLY AND OBSERVED WATER COMING OUT ALONG THE SIDE OF THE BIG BLUE FILTER HOUSING. THE TECHNICIAN FURTHER INSPECTED AND OBSERVED THE FILTER HOUSING WAS CRACKED ALONG ITS LENGTH. THE TECHNICIAN REPLACED THE FILTER HOUSING, TESTING THE UNIT INCLUDING RUNNING A DIAGNOSTIC CYCLE AND FOUND IT TO BE OPERATIONAL. THE TECHNICIAN STATED THAT HE SPOKE WITH THE USER FACILITY BIOMEDICAL DEPARTMENT AND WAS TOLD THAT THEY WERE HAVING AN ISSUE WITH THE MAIN BUILDING PRESSURE REGULATOR. THE CAUSE OF THE EVENT IS ATTRIBUTED TO THE HIGH FACILITY WATER PRESSURE. SUBSEQUENT OF THE EVENT THE USER FACILITY HAS INSTALLED A PRESSURE REGULATOR BEFORE THE SYSTEM 1E. THE BIG BLUE FILTER IS NOT A STERIS PRODUCT, WE DO NOT MANUFACTURE THIS PRODUCT. THIS IS NOT A PRODUCT MARKETED OR PLACE INTO INTERSTATE COMMERCE BY STERIS. THE BIG BLUE FILTERS ARE NECESSARY BASED ON THE USER INCOMING WATER PRESSURE AND ARE THE RESPONSIBILITY OF THE USER'S TO MAINTAIN.
THE USER FACILITY REPORTED THAT THEIR BIG BLUE FILTER CRACKED ALLOWING WATER TO FLOW INTO THE OR, ADJACENT OR ROOMS AND INTO THE BASEMENT SPD. NO INJURY OR PROCEDURAL DELAYS/CANCELLATIONS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STERIS SYSTEM 1E | LIQUID CHEMICAL STERILANT PROCESSING SYSTEM | MED | STERIS CORPORATION - HOPKINS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |