ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2012-14867
- Event Type
- Malfunction
- Date Received
- December 19, 2012
- Report Date
- November 27, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION 01/16/2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE FULLY INTACT; NO PEELING OR DAMAGE WAS OBSERVED. DURING TESTING, THE UP ARROW KEYPAD BUTTON WAS FOUND TO BE INTERMITTENTLY UNRESPONSIVE; THE DOWN ARROW, OK AND CONTRAST KEYPAD BUTTONS RESPONDED APPROPRIATELY AND HAVE SPRING BACK AND CLICK. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER THE UP ARROW, DOWN ARROW AND CONTRAST BUTTON CONTACTS.
ON (B)(6) 2012, THE REPORTER CONTACTED ANIMAS, STATING THAT THE UP ARROW, DOWN ARROW, AND OK KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |