FDA Adverse Event
Injury
Summary report: N
TURBOHAWK ® PLAQUE EXCISION SYSTEM
MDR report key: 2880077
·
Received December 19, 2012
Report
- Report Number
- 2183870-2012-00240
- Event Type
- Injury
- Date Received
- December 19, 2012
- Date of Event
- September 10, 2012
- Report Date
- December 6, 2012
- Manufacturer
- EV3 INC.
- Product Code
- MCW
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
A REVIEW OF THE MANUFACTURE RECORDS FOR THIS DEVICE DID NOT REVEAL ANY DISCREPANCIES RELEVANT TO THE REPORTED EVENT.
Description of Event or Problem · 1
THIS PROCEDURE WAS PERFORMED IN (B)(6) AND IS PART OF (B)(6). THE PROCEDURE DATE WAS (B)(6) 2012. A DISTAL EMBOLISM (PLAQUE, THROMBUS, OR DEBRIS) WAS REPORTED. PTA WAS PERFORMED WITHIN THE STUDY PROCEDURE OF THE PERONEAL ARTERY. DESPITE THE USE OF THE SPIDERFX FILTER, POST PROCEDURAL OCCLUSION OF THE PERONEAL ARTERY OCCURRED DUE TO SMALL EMBOLUS. REVASCULARIZATION BY BALLOON ANGIOPLASTY WAS PERFORMED. PLEASE REFERENCE MDR 2183870-2012-00241 FOR THE SPIDERFX USED IN THIS PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TURBOHAWK ® PLAQUE EXCISION SYSTEM | CATHETER, PERIPHERAL, ATHERECTOMY | MCW | EV3 INC. | THS-LX-C | 9618135 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Required Intervention | COTAVANCE BALLOON| AVANT CORDIS 7F SHEATH| SPIDERFX |