FDA Adverse Event Malfunction Summary report: N

PROPLEGE CORONARY SINUS CATHETER

MDR report key: 2880068 · Received December 19, 2012

Report

Report Number
3008500478-2012-00347
Event Type
Malfunction
Date Received
December 19, 2012
Date of Event
December 6, 2012
Report Date
February 1, 2013
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DWF
PMA / PMN Number
5120780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: THE CATHETER WAS VISUALLY INSPECTED AND FOUND THE TIP OF THE CATHETER WAS STUCK INSIDE THE CONTAMINATION GUARD CAP OF THE CATHETER NO DEVICE DEFECTS WERE CONFIRMED. THE ROOT CAUSE OF THE DIFFICULTIES EXPERIENCED IN THIS CASE CAN MOST LIKELY BE ATTRIBUTED TO THE USER KINKING THE TIP OF THE DEVICE DURING USE. CAPA NOT REQUIRED PER ASSESSMENT. NO CORRECTIVE ACTIONS DUE TO THIS COMPLAINT. THE INFORMATION GATHERED DURING THIS REPORTED EVENT AND EVALUATION WILL BE INCLUDED IN TRENDING DATA FOR FUTURE CAPA DETERMINATIONS. INSTRUCTIONS FOR USE WERE REVIEWED AND FOUND ACCEPTABLE. THE LOT NUMBER OF THE DEVICE IN QUESTION IS UNKNOWN; THEREFORE, A DHR REVIEW CANNOT BE COMPLETED. FROM (B)(4) 2012 (PROPLEGE LAUNCH) TO DATE, THERE HAVE BEEN NO OTHER COMPLAINTS REGARDING THE BALLOON OF THE PROPLEGE DEVICE BEING ECCENTRIC OR HAVING FOREIGN MATTER ATTACHED TO IT. (B)(4). TRENDS WILL CONTINUE TO BE MONITORED ON A MONTHLY BASIS AND IF FURTHER ACTION IS REQUIRED, APPROPRIATE INVESTIGATION WILL BE PERFORMED.

Additional Manufacturer Narrative · 1

DEVICE IS CURRENTLY UNDER EVALUATION INTO ROOT CAUSE.

Description of Event or Problem · 1

IT WAS REPORTED BY THE SALES REP THAT THE CARDIOPLEGIA WAS NOT BEING DELIVERED TO ARREST THE HEART. PER THE REP IT APPEARED AS THE BALLOON OF THE PROPLEGE CORONARY SINUS CATHETER WAS NOT COMPLETELY ROUND AND THERE WAS A "PIECE SEWN TO THE SIDE" OF THE BALLOON. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROPLEGE CORONARY SINUS CATHETER CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS DWF EDWARDS LIFESCIENCES PR9

Patients

Seq Age Sex Outcome Treatment
1