FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 2880043 · Received December 19, 2012

Report

Report Number
2955842-2012-01365
Event Type
Malfunction
Date Received
December 19, 2012
Date of Event
November 20, 2012
Report Date
November 20, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DOES NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY, PROCEDURE A CABLE ON THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT BROKE. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED WITH AN INSTRUMENT OF THE SAME TYPE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10121018 412

Patients

Seq Age Sex Outcome Treatment
1 34 YR DA VINCI SI SURGICAL SYSTEM AND ACCESSORIES