XCEL DILATING TIP TROCAR 5MM
Report
- Report Number
- 3005075853-2012-05762
- Event Type
- Malfunction
- Date Received
- December 17, 2012
- Date of Event
- December 4, 2012
- Report Date
- December 7, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GCJ
- PMA / PMN Number
- K032676
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). WERE ANY NOISES HEARD SUCH AS WHISTLING OR HISSING? BOO. IF SO, DID THE NOISE PREVENT INSUFFLATION?---YES. WAS THERE A DROP IN PRESSURE? IF YES, DID THIS AFFECT THE VISIBILITY OF THE SURGEON? ---NO INFORMATION . WHAT WAS THE GAS CONSUMPTION RATE OR VOLUME (LITER/MINUTE)? ---NO INFORMATION. WERE YOU ABLE TO IDENTIFY WHERE THE LEAK WAS COMING FROM? ---VALVE. WAS A DEVICE INSERTED IN THE TROCAR DURING THE LEAKING? ---NO INFORMATION. IF SO, WHAT DEVICE? WAS ANY TORQUE BEING APPLIED TO THE TROCAR OR DEVICE? ---NO INFORMATION. THE ANALYSIS RESULTS FOUND THAT DEVICES (A AND B) WERE RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICES WERE FUNCTIONALLY TESTED TO DETECT ANY LEAKING ISSUES. A LEAK TEST WAS PERFORMED AND THE DEVICES WERE NOTED TO LEAK DURING INSERTION AND REMOVAL OF THE TEST PROBE THROUGH EACH DEVICE. IT SHOULD BE NOTED THAT AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE POTENTIAL ROOT CAUSE OF THE REPORTED INSUFFLATIONS ISSUES. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4).
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, AIR LEAKED AND A BOO SOUND CONTINUED TO BE HEARD DURING USE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | XCEL DILATING TIP TROCAR 5MM | LAPAROSCOPE, GENERAL AND PLASTIC SURGERY | GCJ | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |