FDA Adverse Event Malfunction Summary report: N

OT ULTRALINK METER

MDR report key: 2874927 · Received December 16, 2012

Report

Report Number
3008382007-2012-08248
Event Type
Malfunction
Date Received
December 16, 2012
Report Date
November 30, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K073231
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING INACCURATE RESULTS. THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE A PATIENT CONTACTED LIFESCAN ALLEGING THAT THE SUBJECT METER READ INACCURATELY HIGH COMPARED TO THE ANOTHER METER. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "385 MG/DL" WITH A LIFESCAN METER AND "154 MG/DL" ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF <= 30% AND/OR <= 30 MG/DL. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT ULTRALINK METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL

Patients

Seq Age Sex Outcome Treatment
1