FDA Adverse Event Malfunction Summary report: N

ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM

MDR report key: 2874872 · Received December 16, 2012

Report

Report Number
2024168-2012-07990
Event Type
Malfunction
Date Received
December 16, 2012
Date of Event
December 3, 2012
Report Date
December 4, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
NIO
PMA / PMN Number
P110028
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN THE ABSOLUTE PRO DEVICE WAS REMOVED FROM THE PACKAGING, DURING PREPARATION, THE STENT IMPLANT WAS OBSERVED TO BE PARTIALLY DEPLOYED. THE DEVICE WAS NOT USED ON THE PATIENT. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. A NEW ABSOLUTE PRO WAS USED TO COMPLETE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM SELF EXPANDING STENT SYSTEM NIO AV-TEMECULA-CT 2051661

Patients

Seq Age Sex Outcome Treatment
1