FDA Adverse Event
Malfunction
Summary report: N
CONTOUR TS
MDR report key: 2874513
·
Received December 14, 2012
Report
- Report Number
- 1826988-2012-00657
- Event Type
- Malfunction
- Date Received
- December 14, 2012
- Date of Event
- November 1, 2012
- Report Date
- November 28, 2012
- Manufacturer
- BAYER HEALTHCARE LLC
- Product Code
- NBW
- PMA / PMN Number
- K062058
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER TESTED HER BLOOD GLUCOSE USING HER CONTOUR TS METER AND RECEIVED READINGS OF 215 AND 223 MG/DL. SHE RETESTED USING ANOTHER METER AND RECEIVED A READING OF 118MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVALUATION. REPLACEMENT TEST STRIPS AND A NEW METER WERE SENT TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTOUR TS | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE LLC | 1820 | 2ED3E04 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |