FDA Adverse Event
Malfunction
Summary report: N
CONTOUR
MDR report key: 2874505
·
Received December 14, 2012
Report
- Report Number
- 1826988-2012-00649
- Event Type
- Malfunction
- Date Received
- December 14, 2012
- Date of Event
- November 1, 2012
- Report Date
- November 20, 2012
- Manufacturer
- BAYER HEALTHCARE LLC
- Product Code
- NBW
- PMA / PMN Number
- K091820
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER TESTED HIS BLOOD GLUCOSE USING HIS CONTOUR USB METER AND RECEIVED A READING OF 200MG/DL. HE RETESTED USING ANOTHER METER AND RECEIVED A READING OF 54MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "D" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVALUATION. REPLACEMENT TEST STRIPS AND A NEW METER WERE SENT TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTOUR | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE LLC | 7098B | 2EC3D05 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |