FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2874498 · Received December 14, 2012

Report

Report Number
2531779-2012-14599
Event Type
Malfunction
Date Received
December 14, 2012
Report Date
November 14, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 (B)(4) 2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH WAS DUPLICATED DURING TESTING. DURING THE PUMP LOAD CARTRIDGE STEP, THE PUMP FAILED TO DETECT THE FILLED CARTRIDGE. A FORCE SENSOR CALIBRATION TEST WAS PERFORMED AND THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. THE PUMP WAS OPENED AND CONTAMINATION WAS FOUND IN THE FORCE SENSOR ASSEMBLY. UNRELATED TO THE COMPLAINT, THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENT AND CONTAMINATION WAS FOUND UNDER THE BUTTON CONTACTS. ALSO UNRELATED, THE BOLUS BUTTON WAS FOUND TO BE TORN AND WAS HARD TO PRESS. ALSO UNRELATED, THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED. THE USER GUIDE INSTRUCTS THE PATIENT THAT CRACKS, CHIPS, OR DAMAGE TO THE PUMP MAY IMPACT THE BATTERY CONTACT AND / OR THE WATERPROOF FEATURE OF THE PUMP. ALSO UNRELATED, THE DISPLAY SCREEN WAS FOUND TO BE FADED. PLEASE DISREGARD REPORT NUMBER 2531779-2012-14415 WHICH IS A DUPLICATE OF THIS REPORT.

Description of Event or Problem · 1

ON (B)(4) 2012, THE PATIENT CONTACTED ANIMAS AND REPORTED A LOAD STEP MALFUNCTION. THE PUMP REPORTEDLY DISPENSED INSULIN DURING THE LOAD STEP EMPTYING THE CARTRIDGE. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED DURING TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 46 YR