FDA Adverse Event Injury Summary report: N

GALAXY INTRAVASCULAR ULTRASOUND SYSTEM

MDR report key: 2874370 · Received December 14, 2012

Report

Report Number
2134265-2012-08329
Event Type
Injury
Date Received
December 14, 2012
Date of Event
October 17, 2012
Report Date
December 13, 2012
Manufacturer
BOSTON SCIENTIFIC - FREMONT (SUD)
Product Code
IYO
PMA / PMN Number
K980851
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GB
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE ROOT CAUSE IS ANTICIPATED PROCEDURAL COMPLICATIONS AS THIS EVENT IS A KNOWN PHYSIOLOGICAL EFFECT OF THE PROCEDURE AND IS NOTED WITHIN THE DFU. (B)(4).

Additional Manufacturer Narrative · 1

UPDATED: PATIENT SEX, RELEVANT TESTS/LAB DATA; OTHER RELEVANT HISTORY. UPDATED/CORRECTION: DESCRIBE EVENT OR PROBLEM. (B)(4).

Description of Event or Problem · 1

SAME CASE AS MDR ID#: 2134265-2012-06875, 2134265-2012-07357. IT WAS REPORTED THAT DURING AN INTRAVASCULAR ULTRASOUND DIAGNOSIS (IVUS) PROCEDURE, CATHETER PULL BACK DIFFICULTIES OCCURRED AND THROMBUS FORMED IN THE TARGET VESSEL. VASCULAR ACCESS WAS OBTAINED VIA RADIAL ARTERY. THE TARGET LESION WAS LOCATED IN THE OSTIAL LEFT CIRCUMFLEX ARTERY (LCX). INITIALLY THE ILAB CART MOTORDRIVE AND THE ATLANTIS SR PRO IMAGING CATHETER FAILED TO PERFORM AN AUTOMATIC PULLPACK, BUT SUCCESSFULLY DID IT ON A THIRD ATTEMPT. AT THAT TIME THROMBUS WAS FOUND IN THE TARGET VESSEL, WHICH WAS SUCCESSFULLY EXTRACTED BY AN ASPIRATION CATHETER. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO PATIENT HARM. THE DRUG THERAPY USED DURING THE PROCEDURE WAS UNKNOWN. THE PHYSICIAN FELT THE EVENT IS CONSIDERED AS LIKELY TO BE RELATED TO THE ATLANTIS IMAGING CATHETER AND NOT THE ILAB PULLBACK FAILURE.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE PATIENT WAS ADMITTED WITH AN EPISODE OF CENTRAL CHEST DISCOMFORT. THE PATIENT DEVELOPED INFEROLATERAL ST SEGMENT DEPRESSION WITH NO SYMPTOMS. CARDIAC CATHETERIZATION WAS THEREFORE RECOMMENDED. INTRAVASCULAR ULTRASOUND CONFIRMED THAT THE MID LAD DISEASE WAS SIGNIFICANT, AND THE OSTIAL DISEASE WAS ALSO SEVERE WITH A VERY HEAVY PLAQUE BURDEN EXTENDING RIGHT BACK TO THE LEFT MAIN STEM. ANGIOGRAPHY REVEALED A FILLING DEFECT IN THE MID LAD ASSUMED TO BE A THROMBUS. HEPARIN 3,000 UNITS OVER AND ABOVE THE ORIGINAL 7,000 UNITS WAS GIVEN ALTHOUGH AN ACT CAME BACK AT 240. ABCIXIMAB WAS GIVEN AS A BOLUS AND INFUSION AND THREE EXPORT RUNS WERE UNDERTAKEN TO ASPIRATE THE CLOT. AT THE END OF THESE PROCEDURES THE LAD LOOKED SATISFACTORY WITH NO OBVIOUS THROMBUS. THE PATIENT WAS COMPLETELY ASYMPTOMATIC AND WAS TRANSFERRED BACK TO CCU. AT THIS STAGE THE PATIENT WAS RECOMMENDED FOR CORONARY ARTERY BYPASS GRAFTING PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GALAXY INTRAVASCULAR ULTRASOUND SYSTEM SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC IYO BOSTON SCIENTIFIC - FREMONT (SUD) UNK428

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention ILAB CART, MDU| ATLANTIS IMAGING CATHETER