Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED A PHOTO OF THE SAMPLE FOR EVALUATION. PHOTOGRAPHIC EVALUATION CONFIRMED THE REPORTED CONDITION OF DEVICE THAT LEAKED DURING FILLING. HOWEVER, DUE TO SAMPLE UNAVAILABILITY, A LEAK TEST COULD NOT BE PERFORMED AND THE ROOT CAUSE COULD NOT BE IDENTIFIED. NO ADDITIONAL OBSERVATION WAS NOTED FROM THE PHOTO. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. ADDITIONAL INFORMATION: THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). THE SAMPLE WAS DISCARDED DUE TO CONTAMINATION. HOWEVER, PICTURES OF THE SAMPLE WERE PROVIDED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT OF AN INFUSOR THAT LEAKED DURING FILLING. THE CONCOMITANT MEDICAL PRODUCTS ARE CURRENTLY UNKNOWN. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 11J066