ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2012-14586
- Event Type
- Malfunction
- Date Received
- December 14, 2012
- Report Date
- November 20, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE TORN OVER THE UP ARROW KEYPAD BUTTON. A DAMAGED KEYPAD WILL ALLOW CONTAMINATION TO PERMEATE THE BUTTON CONTACTS NEGATIVELY IMPACTING THE BUTTON FUNCTION. THE DAMAGED KEYPAD IS OBVIOUS AND DETECTABLE AND SHOULD ALERT THE USER TO DISCONTINUE USE OF THE PUMP. THE UP ARROW AND OK KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE DURING TESTING. THE DOWN AND CONTRAST KEYPAD BUTTONS RESPONDED APPROPRIATELY. DURING INVESTIGATION, EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL KEYPAD CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(4) 2012, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE UP ARROW, DOWN ARROW AND OK KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |