FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX¿

MDR report key: 2874105 · Received December 14, 2012

Report

Report Number
2134265-2012-08083
Event Type
Malfunction
Date Received
December 14, 2012
Date of Event
December 4, 2012
Report Date
December 5, 2012
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR, EVAL SUMMARY ATTACHED, METHOD CODES, RESULT CODES, CONCLUSION CODES: UPDATED. DEVICE EVALUATED BY MFR: THERE WAS BLOOD AND CONTRAST IN THE BALLOON AND INFLATION LUMEN. MAGNIFIED INSPECTION REVEALED NO DAMAGE. WHEN POSITIVE PRESSURE WAS APPLIED WITH AN INFLATION DEVICE FILLED WITH WATER, A STREAM OF WATER EMITTED FROM A PINHOLE IN THE BALLOON WALL ALIGNED WITH THE DISTAL MARKERBAND. THE BALLOON PRESENTED NO IRREGULARITIES IN THE BALLOON MATERIAL OR THE RO MARKER THAT COULD HAVE CONTRIBUTED TO THE DAMAGE. THERE WAS NO EVIDENCE OF ANY PRODUCT QUALITY DEFICIENCIES CONTRIBUTING TO THE DAMAGE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION A BALLOON RUPTURE OCCURRED. THE 90% STENOSED LESION BEING TREATED WAS LOCATED IN THE MODERATELY TORTUOUS DISTAL RIGHT CORONARY ARTERY. THE PHYSICIAN ADVANCED A MM X 2.25MM NC QUANTUM APEX MR BALLOON CATHETER TO THE TARGET LESION. UPON THE FIRST INFLATION AT 18ATMS THE BALLOON RUPTURED. THE DEVICE WAS SUCCESSFULLY REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION A BALLOON RUPTURE OCCURRED. THE 90% STENOSED LESION BEING TREATED WAS LOCATED IN THE MODERATELY TORTUOUS DISTAL RIGHT CORONARY ARTERY. THE PHYSICIAN ADVANCED A MM X 2.25MM NC QUANTUM APEX MR BALLOON CATHETER TO THE TARGET LESION. UPON THE FIRST INFLATION AT 18ATMS THE BALLOON RUPTURED. THE DEVICE WAS SUCCESSFULLY REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NC QUANTUM APEX¿ CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912420220 14834504

Patients

Seq Age Sex Outcome Treatment
1