FDA Adverse Event Malfunction Summary report: N

LIBERTÉ¿

MDR report key: 2874092 · Received December 14, 2012

Report

Report Number
2134265-2012-08081
Event Type
Malfunction
Date Received
December 14, 2012
Date of Event
December 5, 2012
Report Date
December 5, 2012
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
MAF
PMA / PMN Number
P040016
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER - A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND THAT THE STENT WAS PULLED DISTALLY ON THE BALLOON. AS A RESULT THE PROXIMAL EDGE OF THE STENT WAS POSITIONED AT 1CM DISTAL TO THE DISTAL EDGE OF THE PROXIMAL MARKERBAND. THE STENT WAS BUNCHED IN ITS MID- SECTION. THE BALLOON DID NOT APPEAR TO HAVE BEEN SUBJECTED TO ANY POSITIVE PRESSURES. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED HANDLING DAMAGE AS THE EVENT OCCURRED PRIOR TO PATIENT CONTACT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO A PERCUTANEOUS CORONARY INTERVENTION, STENT DAMAGE OCCURRED. AS THE BALLOON PROTECTOR ON THE 35X32MM LIBERTE STENT WAS REMOVED SLIGHT RESISTANCE WAS NOTED, AND UPON REMOVAL IT WAS NOTED THAT THE MIDDLE OF THE STENT WAS BUNCHED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. AS THERE WAS NO CONTACT WITH THE PATIENT, NO COMPLICATIONS WERE REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO A PERCUTANEOUS CORONARY INTERVENTION, STENT DAMAGE OCCURRED. AS THE BALLOON PROTECTOR ON THE 35X32MM LIBERTE STENT WAS REMOVED SLIGHT RESISTANCE WAS NOTED, AND UPON REMOVAL IT WAS NOTED THAT THE MIDDLE OF THE STENT WAS BUNCHED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. AS THERE WAS NO CONTACT WITH THE PATIENT, NO COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIBERTÉ¿ STENT, CORONARY MAF BOSTON SCIENTIFIC - GALWAY H7493893832350 0013290525

Patients

Seq Age Sex Outcome Treatment
1