CLEARLINK SYSTEM CONTINU-FLO4 WAY STOPCOCK EXT SET
Report
- Report Number
- 1416980-2012-07364
- Event Type
- Malfunction
- Date Received
- December 14, 2012
- Date of Event
- November 1, 2012
- Report Date
- November 19, 2012
- Manufacturer
- BAXTER HEALTHCARE - AIBONITO
- Product Code
- FMG
- PMA / PMN Number
- K003225
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NURSE
Narratives
(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. THE LOT NUMBER IS KNOWN AND A BATCH REVIEW WILL BE PERFORMED. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). EVALUATION SUMMARY: THE ACTUAL SAMPLE WAS RECEIVED FOR EVALUATION. THE CUSTOMER REPORTED CONDITION WAS CONFIRMED DURING THE VISUAL INSPECTION. THE ROOT CAUSE WAS NOT IDENTIFIED. THIS ISSUE HAS BEEN ESCALATED TO CAPA. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT.
AN EMAIL WAS RECEIVED FROM A CLINICAL SERVICES MANAGER (CSM) WHO STATED THAT A REGISTERED NURSE (RN) REPORTED HAVING PROBLEMS WITH THE STOP COCK CONNECTIONS IN THEIR ANESTHESIA SET COMING APART AND LEAKING. THE NURSING STAFFS ARE TIGHTENING THE CONNECTIONS PRIOR TO PRIMING THE SET. LOT UR12H16062 IS FROM ONE OF THE SETS. RN IS CONCERNED ABOUT SAFETY AS THEY COULD END UP LOSING THE IV SITES. THE CSM HAVE PERFORMED FOLLOW UP AND REINFORCED THE NEED TO TIGHTEN THE CONNECTIONS PRIOR TO PRIMING THE TUBING. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO ADDITIONAL INFORMATION WAS AVAILABLE. THE RN LEFT A MESSAGE FOR PRODUCT SURVEILLANCE ON (B)(6) 2012 INDICATING THAT SHE SPOKE WITH HER STAFF AGAIN TODAY AND THEY ARE CONTINUING TO FIND LEAKING AT BOTH SIDES OF THE STOP COCK (DISTAL AND PROXIMAL ENDS). SHE SAID THE STAFF IS RE-TIGHTENING THEM AND SOME ARE STILL LEAKING AND HAVE TO TIGHTEN THEM EVEN MORE. SHE SAID THE FRUSTRATION WAS HAVING TO TIGHTEN EACH SET SINCE THEY COME LOOSE FROM THE PACKAGE AND THEY HAVEN'T HAD THAT PROBLEM UNTIL RECENTLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CLEARLINK SYSTEM CONTINU-FLO4 WAY STOPCOCK EXT SET | STOPCOCK, I.V. SET | FMG | BAXTER HEALTHCARE - AIBONITO | UR12H16062 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |