FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 2874056 · Received December 14, 2012

Report

Report Number
2531779-2012-14574
Event Type
Injury
Date Received
December 14, 2012
Date of Event
November 15, 2012
Report Date
November 15, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION (B)(4) 2013 DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A RESERVED SAMPLE FROM THE SAME LOT NUMBER WAS TESTED. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED AND NO DAMAGE OR DEFECTS WERE NOTED. A FILL, FORCE AND LEAK TEST WAS PERFORMED WITH NO FAILURES. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.

Additional Manufacturer Narrative · 1

THE CARTRIDGE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED ANIMAS STATING THAT THERE WERE AIR BUBBLES IN THE CARTRIDGE AND THE PATIENT EXPERIENCED A BLOOD GLUCOSE (BG) VALUE OF 540MG/DL WITH NO SYMPTOMS. THE PATIENT REPORTEDLY WAS TREATED VIA CORRECTION INJECTION. IT WAS NOTED THAT ROOM TEMPERATURE INSULIN WAS NOT USED. CUSTOMER SUPPORT (CS) REVIEWED THE REPORTER'S TECHNIQUE. USE ERROR CONTRIBUTED TO REPORTED EVENT BASED ON THE FOLLOWING REASONS: THE INSULIN VIAL WAS INCORRECTLY PRESSURIZED AND THE INSULIN BEING USED WAS NOT AT ROOM TEMPERATURE. CS ADVISED THE REPORTER TO USE ROOM TEMPERATURE INSULIN PRIOR TO FILLING THE CARTRIDGE. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGATION THAT THE PATIENT EXPERIENCED A HYPERGLYCEMIC EVENT WHILE USING INSULIN PUMP THERAPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM IR1200/1250/2020/OTP CART

Patients

Seq Age Sex Outcome Treatment
1 8 YR Life Threatening