FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX¿

MDR report key: 2873690 · Received December 14, 2012

Report

Report Number
2134265-2012-07784
Event Type
Malfunction
Date Received
December 14, 2012
Date of Event
November 27, 2012
Report Date
November 27, 2012
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: THERE WAS BLOOD IN THE BALLOON AND INFLATION LUMEN. MAGNIFIED INSPECTION REVEALED NO DAMAGE. WHEN POSITIVE PRESSURE WAS APPLIED WITH AN INFLATION DEVICE FILLED WITH WATER, A STREAM OF WATER EMITTED FROM A PINHOLE IN THE BALLOON WALL 2MM FROM THE PROXIMAL EDGE OF THE DISTAL MARKERBAND. THE BALLOON PRESENTED NO IRREGULARITIES IN THE BALLOON MATERIAL OR THE RO MARKER THAT COULD HAVE CONTRIBUTED TO THE DAMAGE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL / PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE, A BALLOON RUPTURE OCCURRED. THE PROCEDURE TREATED THE 99% STENOSED TARGET LESION LOCATED IN THE SEVERELY CALCIFIED AND SEVERELY TORTUOUS MID LEFT ANTERIOR DESCENDING ARTERY. A 3.0 X 8.0 MM NC QUANTUM APEX BALLOON WAS ADVANCED AND USED FOR PRE-DILATION, HOWEVER THE BALLOON RUPTURED DUE TO SEVERE CALCIFICATION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE, A BALLOON RUPTURE OCCURRED. THE PROCEDURE TREATED THE 99% STENOSED TARGET LESION LOCATED IN THE SEVERELY CALCIFIED AND SEVERELY TORTUOUS MID LEFT ANTERIOR DESCENDING ARTERY. A 3.0X8.0MM NC QUANTUM APEX BALLOON WAS ADVANCED AND USED FOR PRE-DILATION, HOWEVER THE BALLOON RUPTURED DUE TO SEVERE CALCIFICATION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NC QUANTUM APEX¿ CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912408300 15223939

Patients

Seq Age Sex Outcome Treatment
1