FDA Adverse Event Malfunction Summary report: N

ENTERALITE INFINITY ENTERAL FEEDING PUMP

MDR report key: 2873496 · Received November 15, 2012

Report

Report Number
1722139-2012-01170
Event Type
Malfunction
Date Received
November 15, 2012
Date of Event
October 1, 2012
Report Date
November 7, 2012
Manufacturer
MOOG MED DEVICES GROUP
Product Code
LZH
PMA / PMN Number
K031199
Removal / Correction Number
TBD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FOUND THAT THE DIODE D7 WAS DETERMINED TO BE LEAKY AND WAS OUT OF THE SPECIFICATION. NEW PCB WAS REPLACED TO SOLVE THE ISSUE.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THAT PUMP FAILED THE DIODE TEST WHICH IS THE CAUSE OF THE ERROR CODE 13 DURING TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTERALITE INFINITY ENTERAL FEEDING PUMP LZH MOOG MED DEVICES GROUP INFINITY PUMP

Patients

Seq Age Sex Outcome Treatment
1