FDA Adverse Event Malfunction Summary report: N

MANUAL ORTHOPEDIC SURGICAL INSTRUMENT

MDR report key: 2873419 · Received December 14, 2012

Report

Report Number
1030489-2012-02811
Event Type
Malfunction
Date Received
December 14, 2012
Date of Event
November 16, 2012
Report Date
November 16, 2012
Manufacturer
MSD BARTLETT MFG
Product Code
LXH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

VISUALLY CONFIRMED APPROXIMATELY ~4MM OF THE INSTRUMENT TIP HAS BEEN BROKEN OFF, CONSISTENT WITH INTERFACE DURING USAGE. INSPECTION OF THE SHAFT DIAMETER AND MATERIAL HARDNESS CONFIRMED CONFORMANCE TO PRINT SPECIFICATION. FRACTURE SURFACE ANALYSIS OF BOTH INSTRUMENTS IDENTIFIED A FAIRLY FLAT FRACTURE SURFACE AND CIRCULAR MATERIAL MOVEMENT, CONSISTENT WITH TORSIONAL OVERLOAD. THE ABOVE FINDINGS ARE CONSISTENT WITH TORSIONAL OVERLOAD.

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

(B)(4): NEITHER THE DEVICE NOR APPLICABLE IMAGING FILMS WERE RETURNED TO THE MANUFACTURER FOR EVALUATION, THEREFORE, THE CAUSE OF THE EVENT CANNOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A POSTERIOR SURGICAL PROCEDURE AT L5-S. ON 7-8 YEARS POST-OP THE PATIENT UNDERWENT A REMOVAL SURGERY DUE TO COMPLETION OF BONE UNION. DURING REMOVAL OF SCREWS, THE TIP OF THE DRIVER BROKE OFF. THE FRAGMENT WAS STUCK IN THE SCREW POST. THE SURGEON WAS ABLE TO REMOVE THE SCREW USING ANOTHER INSTRUMENT SUCCESSFULLY. NO FRAGMENT WAS LEFT IN THE PATIENT. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MANUAL ORTHOPEDIC SURGICAL INSTRUMENT ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH MSD BARTLETT MFG NA JM04D030

Patients

Seq Age Sex Outcome Treatment
1 00051 YR