MANUAL ORTHOPEDIC SURGICAL INSTRUMENT
Report
- Report Number
- 1030489-2012-02811
- Event Type
- Malfunction
- Date Received
- December 14, 2012
- Date of Event
- November 16, 2012
- Report Date
- November 16, 2012
- Manufacturer
- MSD BARTLETT MFG
- Product Code
- LXH
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
VISUALLY CONFIRMED APPROXIMATELY ~4MM OF THE INSTRUMENT TIP HAS BEEN BROKEN OFF, CONSISTENT WITH INTERFACE DURING USAGE. INSPECTION OF THE SHAFT DIAMETER AND MATERIAL HARDNESS CONFIRMED CONFORMANCE TO PRINT SPECIFICATION. FRACTURE SURFACE ANALYSIS OF BOTH INSTRUMENTS IDENTIFIED A FAIRLY FLAT FRACTURE SURFACE AND CIRCULAR MATERIAL MOVEMENT, CONSISTENT WITH TORSIONAL OVERLOAD. THE ABOVE FINDINGS ARE CONSISTENT WITH TORSIONAL OVERLOAD.
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
(B)(4): NEITHER THE DEVICE NOR APPLICABLE IMAGING FILMS WERE RETURNED TO THE MANUFACTURER FOR EVALUATION, THEREFORE, THE CAUSE OF THE EVENT CANNOT BE DETERMINED.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A POSTERIOR SURGICAL PROCEDURE AT L5-S. ON 7-8 YEARS POST-OP THE PATIENT UNDERWENT A REMOVAL SURGERY DUE TO COMPLETION OF BONE UNION. DURING REMOVAL OF SCREWS, THE TIP OF THE DRIVER BROKE OFF. THE FRAGMENT WAS STUCK IN THE SCREW POST. THE SURGEON WAS ABLE TO REMOVE THE SCREW USING ANOTHER INSTRUMENT SUCCESSFULLY. NO FRAGMENT WAS LEFT IN THE PATIENT. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MANUAL ORTHOPEDIC SURGICAL INSTRUMENT | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | MSD BARTLETT MFG | NA | JM04D030 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00051 YR |