FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2873349 · Received December 14, 2012

Report

Report Number
1416980-2012-07346
Event Type
Malfunction
Date Received
December 14, 2012
Date of Event
December 9, 2012
Report Date
December 9, 2012
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) IS CONFIRMED; PER THE COMPLAINT INFORMATION THE LIKELY CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE BECAUSE THERE WAS A LOOSE CONNECTION BETWEEN LINE AND SUPPLY BAG, WHICH IS A USE ERROR THAT CAN CAUSE SYSTEM ERROR 2240 ALARMS. THE CARE GIVER STATED THE HEATER BAG WAS NOT CONNECTED PROPERLY. A LABELING REVIEW FOUND THE LABELING TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.

Additional Manufacturer Narrative · 1

(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REPORTING A SYSTEM ERROR (SE) 2240/2367 (AIR IN LINE) DURING THE INITIAL DRAIN ON THE HOME CHOICE (HC). THE CAREGIVER (CG) CYCLED THE POWER ON THE HC AND THE HC ALARMED SYSTEM ERROR 2367. THE CG CYCLED THE POWER ON THE HC AND PROCEEDED TO PRESS GO TO START. THE CG STATED THE HEATER BAG WAS NOT CONNECTED PROPERLY. THE TSR ADVISED THE CG TO START OVER WITH ALL NEW SUPPLIES AND INFORM THE REGISTERED NURSE (RN) OF THE SE 2240. THERE WAS PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 78 YR HOMECHOICE