FDA Adverse Event Malfunction Summary report: N

MARYLAND BIPOLAR FORCEPSINSTRUMENT

MDR report key: 2873214 · Received December 13, 2012

Report

Report Number
2955842-2012-01297
Event Type
Malfunction
Date Received
December 13, 2012
Date of Event
June 20, 2012
Report Date
October 3, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING FOUND THE JAWS OF THE INSTRUMENT ARE MISALIGNED. ONE GRIP IS BENT, CAUSING SIDE TO SIDE MISALIGNMENT OF GRIPS. THERE IS A 0.105 OFFSET AT THE TIPS. ELECTRICAL CONTINUITY PASSED. ADDITIONAL FINDINGS REVEALED SCRATCHES ON THE MAINTUBE. DISTAL END OF MAIN TUBE HAS VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. SUGGESTING THE DAMAGE MAY HAVE BEEN CAUSED BY INSTRUMENT COLLISIONS OR ROUGH HANDLING. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S HYSTERECTOMY SURGICAL PROCEDURE THE JAWS OF THE MARYLAND BIPOLAR FORCEPS WERE MISALIGNED SO THE INSTRUMENT COULD NOT PRESS AND COAGULATE PROPERLY. NOTHING WAS REPORTED HAVING FALLEN INTO THE PATIENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MARYLAND BIPOLAR FORCEPSINSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420172-07 M10111128 759

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SYSTEM INSTRUMENTS AND ACCESSORIES