FREESTYLE LITE
Report
- Report Number
- 2954323-2012-06769
- Event Type
- Injury
- Date Received
- December 13, 2012
- Date of Event
- November 20, 2012
- Report Date
- November 21, 2012
- Product Code
- NBW
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE METER WAS RETURNED AND TESTED WITH RETURNED TEST STRIPS. NO NEW ISSUES WERE OBSERVED. DISASSEMBLED METER AND OBSERVED BLOCKAGE OF A PIECE OF PAPER IN THE STRIP PORT. REMOVED PAPER FROM STRIP PORT. CONTROL SOLUTION TESTING WAS PERFORMED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. DID NOT OBSERVE TDN-3 ISSUE. THE COMPLAINT WAS NOT CONFIRMED.
THE PRODUCTS HAVE BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED. (B)(4).
CUSTOMER REPORTED THAT ON (B)(6) 2012 HE WAS EXPERIENCING UNSPECIFIED "LOW BLOOD GLUCOSE" SYMPTOMS SO HE CALLED HIS SON TO ASSIST WITH A BLOOD TEST. HOWEVER, THE TEST DID NOT START ON HIS ADC BLOOD GLUCOSE METER AFTER A SAMPLE WAS APPLIED TO A TEST STRIP INSERTED INTO IT. IT WAS FURTHER REPORTED THAT DUE TO THIS CUSTOMER EXPERIENCED A LOSS OF CONSCIOUSNESS. NO THIRD-PARTY EMERGENT MEDICAL INTERVENTION WAS REQUIRED. CUSTOMER WAS GIVEN FRUIT PUNCH, SWEET TEA AND AN UNSPECIFIED NUMBER OF GLUCOSE TABLETS TO INGEST BY HIS SON. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 1271509 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Other| R |