ARMADA 14 PTA CATHETER
Report
- Report Number
- 2024168-2012-07925
- Event Type
- Malfunction
- Date Received
- December 13, 2012
- Date of Event
- November 20, 2012
- Report Date
- November 21, 2012
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- LIT
- PMA / PMN Number
- K102705
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE BALLOON RUPTURE WAS ABLE TO BE CONFIRMED. BASED ON A VISUAL AND FUNCTIONAL INSPECTION AND SCANNING ELECTRON MICROSCOPY IMAGING OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.
IT WAS REPORTED THAT DURING A PROCEDURE, A 2.0 X 60/150 ARMADA 0.014 BALLOON DILATATION CATHETER (BDC) WAS ADVANCED, WITH NO NOTED RESISTANCE, TOWARD A LESION LOCATED IN AN UNSPECIFIED BELOW-THE-KNEE VESSEL. THE ARMADA BALLOON RUPTURED DURING THE FIRST INFLATION TO 8 ATMOSPHERES. THE ARMADA WAS WITHDRAWN FROM THE ANATOMY WITH NO NOTED RESISTANCE AND THE PROCEDURE WAS COMPLETED USING A NEW ARMADA BDC. THERE WAS NO REPORTED ADVERSE PATIENT EFFECTS. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARMADA 14 PTA CATHETER | PERIPHERAL DILATATION CATHETER | LIT | AV-TEMECULA-CT | 769163 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR |