FDA Adverse Event Malfunction Summary report: N

ARMADA 14 PTA CATHETER

MDR report key: 2873126 · Received December 13, 2012

Report

Report Number
2024168-2012-07925
Event Type
Malfunction
Date Received
December 13, 2012
Date of Event
November 20, 2012
Report Date
November 21, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
LIT
PMA / PMN Number
K102705
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE BALLOON RUPTURE WAS ABLE TO BE CONFIRMED. BASED ON A VISUAL AND FUNCTIONAL INSPECTION AND SCANNING ELECTRON MICROSCOPY IMAGING OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE, A 2.0 X 60/150 ARMADA 0.014 BALLOON DILATATION CATHETER (BDC) WAS ADVANCED, WITH NO NOTED RESISTANCE, TOWARD A LESION LOCATED IN AN UNSPECIFIED BELOW-THE-KNEE VESSEL. THE ARMADA BALLOON RUPTURED DURING THE FIRST INFLATION TO 8 ATMOSPHERES. THE ARMADA WAS WITHDRAWN FROM THE ANATOMY WITH NO NOTED RESISTANCE AND THE PROCEDURE WAS COMPLETED USING A NEW ARMADA BDC. THERE WAS NO REPORTED ADVERSE PATIENT EFFECTS. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARMADA 14 PTA CATHETER PERIPHERAL DILATATION CATHETER LIT AV-TEMECULA-CT 769163

Patients

Seq Age Sex Outcome Treatment
1 76 YR