FDA Adverse Event Malfunction Summary report: N

SEE H-10

MDR report key: 2873109 · Received December 13, 2012

Report

Report Number
2023826-2012-00982
Event Type
Malfunction
Date Received
December 13, 2012
Date of Event
November 15, 2012
Report Date
November 15, 2012
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P880091
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

BRAND NAME: SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCULAR LENS (ELASTIC). (B)(4). EVALUATION: RESULTS: VISUAL INSPECTION OF THE RETURNED LENS SHOWED THE LENS WAS CUT INTO THREE PIECES AND A PIECE OF THE LENS HAD BEEN TORN OFF AND WAS MISSING. THERE WAS A CLEAR SURGICAL RESIDUE ON THE LENS. CONCLUSION: AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED, WHICH WAS SUBSEQUENTLY CLOSED. THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MANUFACTURING OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. BASED ON THE COMPLAINT HISTORY AND THE EVALUATION OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. CLAIM# (B)(4).

Description of Event or Problem · 1

IT WAS REPORTER THAT THE SURGEON INSERTED THE AA4203TL SILICONE SINGLE PIECE LENS AND THE LENS TORE AS IT WAS INSERTED. THE SURGEON CUT THE LENS AND REMOVED IT FROM THE EYE WITHOUT ANY PATIENT INJURY. ANOTHER SAME MODEL LENS WAS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SEE H-10 INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY AA4203TL N/A

Patients

Seq Age Sex Outcome Treatment
1 MODEL NUMBER MTC-60CFP, LOT NUMBER UNKNOWN| MODEL NUMBER MSI-TM, LOT NUMBER UNKNOWN