COULTER® LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2012-02862
- Event Type
- Malfunction
- Date Received
- December 13, 2012
- Date of Event
- November 15, 2012
- Report Date
- November 16, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K011342
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FIELD SERVICE ENGINEER (FSE) NOTED THE SAMPLE TUBE TO THE FLOWCELL WAS REMOVED. THE FSE CLEANED AND DRIED THE AREA WITH GAUZE AND REPLACED THE SAMPLE TUBING TO THE FLOWCELL. THE FSE VERIFIED SYSTEM OPERATION WITH PASSING SYSTEM STARTUP, LATRON, AND CONTROLS. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE UNIT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS. (B)(4).
THE CUSTOMER REPORTED APPROXIMATELY FIVE (5) MILLILITERS OF CLENZ SOLUTION LEAKED ON THE COUNTER WHEN THE INSTRUMENT WAS IN SHUTDOWN INVOLVING THE COULTER LH 750 HEMATOLOGY ANALYZER. THE CUSTOMER HAD ON PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT IMPACTED. THE CUSTOMER HAS A RISK MANAGEMENT PLAN AT THE FACILITY. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER® LH 750 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |