FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 2873073 · Received December 13, 2012

Report

Report Number
1061932-2012-02862
Event Type
Malfunction
Date Received
December 13, 2012
Date of Event
November 15, 2012
Report Date
November 16, 2012
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKZ
PMA / PMN Number
K011342
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) NOTED THE SAMPLE TUBE TO THE FLOWCELL WAS REMOVED. THE FSE CLEANED AND DRIED THE AREA WITH GAUZE AND REPLACED THE SAMPLE TUBING TO THE FLOWCELL. THE FSE VERIFIED SYSTEM OPERATION WITH PASSING SYSTEM STARTUP, LATRON, AND CONTROLS. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE UNIT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY FIVE (5) MILLILITERS OF CLENZ SOLUTION LEAKED ON THE COUNTER WHEN THE INSTRUMENT WAS IN SHUTDOWN INVOLVING THE COULTER LH 750 HEMATOLOGY ANALYZER. THE CUSTOMER HAD ON PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT IMPACTED. THE CUSTOMER HAS A RISK MANAGEMENT PLAN AT THE FACILITY. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COULTER® LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1