TAXUS¿ LIBERTÉ¿
Report
- Report Number
- 2134265-2012-07817
- Event Type
- Malfunction
- Date Received
- December 13, 2012
- Date of Event
- September 12, 2012
- Report Date
- November 19, 2012
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- P060008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CN
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE IS A COMBINATION PRODUCT (B)(4). DEVICE EVALUATED BY MFR: VISUAL EXAMINATION OF THE RETURNED DEVICE NOTED A SEVERE KINK LOCATED APPROXIMATELY 13.5MM PROXIMAL TO THE EXCHANGE PORT. THE PROXIMAL EDGE OF THE STENT WAS ALSO DAMAGED WITH A NUMBER OF STRUTS BENT BACK DISTALLY. THE BALLOON AND TIP SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. NO FURTHER ISSUES WERE NOTED WITH THE RETURNED DEVICE WHICH COULD HAVE CONTRIBUTED TO THE REPORTED COMPLAINT. THE MOST PROBABLE CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE IS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).
REPORTABLE BASED ON ANALYSIS COMPLETED ON (B)(6) 2012. IT WAS REPORTED THAT DURING A CORONARY ARTERY STENTING TREATMENT PROCEDURE, FAILURE TO CROSS THE LESION OCCURRED. THE TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY (LAD). A 2.25X32MM TAXUS LIBERTE STENT WAS SELECTED AND ADVANCED TO TREAT THE TARGET LESION; HOWEVER, THE DEVICE WAS UNABLE TO CROSS THE LESION. THE DEVICE WAS REMOVED INTACT. THE PROCEDURE WAS COMPLETED WITH ANOTHER WITH SAME DEVICE. THERE WAS NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S CONDITION WAS STABLE. HOWEVER, THE RETURNED DEVICE REVEALED STENT DAMAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TAXUS¿ LIBERTÉ¿ | CORONARY DRUG-ELUTING STENT | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493894032220 | 14661768 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR |