FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2872880 · Received December 13, 2012

Report

Report Number
2531779-2012-14542
Event Type
Malfunction
Date Received
December 13, 2012
Report Date
November 16, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY REVEALED THAT THE PUMP DID NOT DETECT THE CARTRIDGE. INVESTIGATION REVEALED THAT THE FORCE SENSOR WAS OUT OF CALIBRATION. THERE WAS EVIDENCE OF MOISTURE AND CONTAMINATION FOUND ON THE FORCE SENSOR HOUSING AND SHIM.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE PUMP DID NOT DETECT THE CARTRIDGE. THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. THERE WAS EVIDENCE OF MOISTURE AND CONTAMINATION FOUND ON THE FORCE SENSOR HOUSING AND SHIM. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 32 YR