FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2872880
·
Received December 13, 2012
Report
- Report Number
- 2531779-2012-14542
- Event Type
- Malfunction
- Date Received
- December 13, 2012
- Report Date
- November 16, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY REVEALED THAT THE PUMP DID NOT DETECT THE CARTRIDGE. INVESTIGATION REVEALED THAT THE FORCE SENSOR WAS OUT OF CALIBRATION. THERE WAS EVIDENCE OF MOISTURE AND CONTAMINATION FOUND ON THE FORCE SENSOR HOUSING AND SHIM.
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE PUMP DID NOT DETECT THE CARTRIDGE. THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. THERE WAS EVIDENCE OF MOISTURE AND CONTAMINATION FOUND ON THE FORCE SENSOR HOUSING AND SHIM. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2012.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR |