PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2012-07910
- Event Type
- Injury
- Date Received
- December 13, 2012
- Date of Event
- November 13, 2012
- Report Date
- December 3, 2012
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE PLUNGER, SUTURE, LINK, NEEDLES, AND CUFFS WERE NOT RETURNED WITH THE DEVICE AND WITHOUT THE RETURN OF THESE COMPONENTS, THE SCOPE OF THIS INVESTIGATION WAS VERY LIMITED AND THE REPORTED CUFF MISS COULD NOT BE CONFIRMED. THE PROBABLE CAUSE FOR THE REPORTED CUFF MISS COULD NOT BE DETERMINED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 20810J1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |