PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 2872718
·
Received December 13, 2012
Report
- Report Number
- 3004209178-2012-91363
- Event Type
- Malfunction
- Date Received
- December 13, 2012
- Date of Event
- December 1, 2012
- Report Date
- December 1, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCE UNEXPLAINED HIGH BLOOD GLUCOSE OF 369MG/DL. TROUBLESHOOTING WAS PERFORMED. ASSISTED THE CALLER TO RUN A MANUAL PRIME TEST AND THE INSULIN DID EXIT. PERFORMED A HIGH PRESSURE TEST AND THE TEST FAILED TWICE. INSTRUCTED THE CALLER TO CHANGE THE ENTIRE INFUSION SET AND RESERVOIR. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |