FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX¿

MDR report key: 2872227 · Received December 13, 2012

Report

Report Number
2134265-2012-07759
Event Type
Malfunction
Date Received
December 13, 2012
Date of Event
November 28, 2012
Report Date
November 29, 2012
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: THE NC QUANTUM APEX CATHETER WAS RECEIVED INSIDE A CARRIER TUBE (HOOP) WITH NO OTHER DEVICES. MAGNIFIED INSPECTION REVEALED NO DAMAGE. WHEN POSITIVE PRESSURE WAS APPLIED WITH AN INFLATION DEVICE FILLED WITH WATER, A STREAM OF WATER EMITTED FROM A PINHOLE IN THE BALLOON WALL 1MM FROM THE DISTAL EDGE OF THE PROXIMAL MARKERBAND. THE BALLOON PRESENTED NO IRREGULARITIES IN THE BALLOON MATERIAL OR THE RO MARKER THAT COULD HAVE CONTRIBUTED TO THE DAMAGE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTIONAL PROCEDURE A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS DISTAL RIGHT CORONARY ARTERY. THE 2.75 X 12 MM NC QUANTUM APEX MR BALLOON CATHETER WAS INFLATED TO 10-12 ATMS AND RUPTURED ON THE FIRST INFLATION. THE BALLOON CATHETER WAS REMOVED INTACT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS STABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTIONAL PROCEDURE A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS DISTAL RIGHT CORONARY ARTERY. THE 2.75X12MM NC QUANTUM APEX MR BALLOON CATHETER WAS INFLATED TO 10-12 ATMS AND RUPTURED ON THE FIRST INFLATION. THE BALLOON CATHETER WAS REMOVED INTACT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NC QUANTUM APEX¿ CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912412270 15354032

Patients

Seq Age Sex Outcome Treatment
1