SWAN-GANZ POLYMER BLEND TORQUE SUPPORT TRUE SIZE TD CATHETER
Report
- Report Number
- 2015691-2012-18839
- Event Type
- Malfunction
- Date Received
- December 13, 2012
- Date of Event
- November 1, 2012
- Report Date
- November 1, 2012
- Manufacturer
- EDWARDS LIFESCIENCES, PR
- Product Code
- DQO
- PMA / PMN Number
- K915726
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
ONE CATHETER WITH MONOJECT 1.5CC LIMITED VOLUME SYRINGE WAS RETURNED FOR EVALUATION. NO INTRODUCER WAS RETURNED. THE BALLOON DID NOT INFLATE DUE TO AN INTERLUMEN LEAKAGE IN THE BACKFORM BETWEEN THE INFLATION AND DISTAL LUMENS. EXAMINATION OF THE CATHETER FOUND THE DISTAL LUMEN TO BE FULLY OCCLUDED. THE PROXIMAL LUMEN WAS FOUND TO BE PATENT AND DID NOT LEAK. THE BALLOON INFLATED ALTHOUGH IT WAS FOUND TO HAVE WHAT APPEARS TO BE BLOOD ON THE INSIDE SURFACE OF THE LATEX. THE OCCLUSION IN THE DISTAL LUMEN WAS CLEARED USING HOT WATER AND A STYLET WIRE. LEAK TESTING WAS THEN PERFORMED AND THE DISTAL LUMEN WAS FOUND TO LEAK INTO THE INFLATION LUMEN FROM THE INSIDE THE BACKFORM. THIS CONDITION CONFIRMS THE CUSTOMER REPORT OF BALLOON COULD NOT MAINTAIN INFLATION DURING USE. BALLOON INFLATION TEST WAS PERFORMED USING RETURNED SYRINGE WITH 1.5CC AIR. VISUAL EXAMINATIONS WERE PERFORMED UNDER MICROSCOPE AT 10X MAGNIFICATION AND WITH UNAIDED EYE. A REVIEW OF THE MANUFACTURING RECORDS INDICATED THAT THE PRODUCT MET SPECIFICATION AT THE TIME OF RELEASE. THE CUSTOMER REPORT WAS CONFIRMED AS RELATED TO THE MANUFACTURING PROCESS. AN INVESTIGATION IS UNDERWAY TO DETERMINE WHAT ACTIONS ARE NEEDED TO PREVENT RECURRENCE OF THIS TYPE OF COMPLAINT.
IT WAS REPORTED THAT THE BALLOON COULD NOT MAINTAIN INFLATION DURING USE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SWAN-GANZ POLYMER BLEND TORQUE SUPPORT TRUE SIZE TD CATHETER | SWAN-GANZ POLYMER BLEND TORQUE SUPPORT TRUE SIZE TD CATHETER | DQO | EDWARDS LIFESCIENCES, PR | 174HF7 | 59335167 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |