FDA Adverse Event Injury Summary report: N

SUMMIT DUOFIX TAP SZ6 STD OFF

MDR report key: 2872109 · Received December 13, 2012

Report

Report Number
1818910-2012-83662
Event Type
Injury
Date Received
December 13, 2012
Date of Event
December 5, 2012
Report Date
May 22, 2013
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
LPH
PMA / PMN Number
PK011489
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPDATE: (B)(6) 2013 PFS WAS RECEIVED FROM LEGAL, MEDICAL RECORDS WERE RECEIVED FROM LEGAL, AND PART/LOT INFORMATION WAS IDENTIFIED. RECORDS ARE AVAILABLE FOR FURTHER REVIEW. THE INVESTIGATION IS ONGOING. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A PREVIOUS REVIEW OF THE DEVICE HISTORY RECORDS FOR THE 2241089 LOT CODE DID NOT REVEAL ANY RELATED MANUFACTURING DEVIATIONS OR ANOMALIES. A COMPLAINT DATABASE SEARCH FOUND ONE ADDITIONAL COMPLAINT AGAINST THE AT3EM1000 LOT CODE. OSTEOLYSIS AND PAIN WAS NOT ALLEGED IN THE PREVIOUSLY REPORTED COMPLAINT. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE REMAINING PRODUCT AND LOT COMBINATION. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

CORRECTED: EVENT/PROBLEM DESCRIPTION, DATE RECEIVED BY MANUFACTURER. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATIONS SINCE THEIR RELEASE FOR DISTRIBUTION. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE FOR THE DEPUY HIP STEM AS THE PRODUCT AND LOT CODE REQUIRED WAS NOT PROVIDED. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A; REV. C. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT WITH THE INFORMATION AVAILABLE. BASED ON THE INABILITY TO IDENTIFY A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS OSTEOLYSIS AND PAIN. IT ALSO APPEARED THAT THE STEM SUBSIDED AND THEN BECAME SOLIDLY FIXED. UPDATE - (B)(4) 2013 - LITIGATION PAPERS RECEIVED. IT IS ALLEGED THAT THE PATIENT SUFFERS FROM PAIN, DISCOMFORT, AND WEAKNESS. THERE IS NO NEW ADDITIONAL INFORMATION THAT WOULD AFFECT THE INVESTIGATION.

Description of Event or Problem · 1

UPDATE REC'D 05/22/2013 - PFS AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF THE PRIMARY AND REVISION OPERATIVE NOTE IT WAS NOTED THAT THE SCREWS WERE REMOVED FOR THE CUP AS THE SURGEON THOUGHT THE CUP WAS LOOSE, BUT THE CUP AND SCREWS ENDED UP BEING WELL FIXED. A LEG LENGTH DISCREPANCY WAS ALSO NOTED. THERE IS NO NEW ADDITIONAL INFORMATION THAT WOULD AFFECT THE EXISTING MDR DECISION.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS OSTEOLYSIS AND PAIN. IT ALSO APPEARED THAT THE STEM SUBSIDED AND THEN BECAME SOLIDLY FIXED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SUMMIT DUOFIX TAP SZ6 STD OFF HIP STEM LPH DEPUY ORTHOPAEDICS INC US AT3EM1000

Patients

Seq Age Sex Outcome Treatment
1 61 YR Required Intervention